55
Note: When adding prismatic power, only add one eye’s prismatic power instead of two eyes’ prismatic powers.
5.5.5 Horizontal coincidence test (horizontal image inequality and horizontal heterophoria)
Test purpose: to test horizontal image inequality and horizontal heterophoria
Test sighting mark: horizontal coincidence sighting mark
Assistant lens: Binocular rotatory prime
Right eye red optical filter, left eye green optical filter (red and green horizontal coincidence
sighting mark)
Binocular circular polarized optical filter (polarized horizontal coincidence sighting mark)
1. Call in horizontal coincidence sighting mark.
2. Ask the patient about the sighting mark he sees and test according his answers.
2.1 Horizontal image inequality
If the sizes of the upper and lower frames are equivalent, there is no horizontal image inequality; if
the sizes of the upper and lower frames are inequivalent, there is horizontal image inequality.
As shown in right figure: about 3.5 image inequality.
The straight line width of the sighting mark amounts to 3.5% image inequality.
Horizontal heterophoria
Sighting
mark shape
Diagnosis Correction
No heterophoria
Align eye position, no need for correction.
Esophoria
Increase BO prismatic power until it the upper frame
aligns at the lower frame.
Exophoria
Increase BI prismatic power until it the upper frame
aligns at the lower frame.
Note: When adding prismatic power, only add one eye’s prismatic power instead of two eyes’ prismatic powers.
5.5.6 Vertical coincidence test (vertical image inequality and vertical heterophoria)
Содержание YPA-2100
Страница 1: ...YPA 2100 DIGITAL REFRACTOR User Manual Version 1 0 Date 20180521...
Страница 5: ...13 Electrical Schematic Diagram 82 14 Electromagnetic Compatibility 83...
Страница 24: ...19 Press to enlarge PD press to reduce PD 3 Press to exit PD input interface...
Страница 31: ...26 key...
Страница 42: ...37 4 Click key shift to 0 5XC 5 Click the key again shift to automatic mode Slit Prism...
Страница 55: ...50 34 Press to finish the test...
Страница 72: ...67 3 Program 1 A B C D are recommended vision test steps users can also add or delete as needed...
Страница 74: ...69 Add Key Click it to add one step automatically 4 Click key to enter program customizing interface...
Страница 78: ...73 3 Click key to popup input box...
Страница 79: ...74 4 Use CCQ 800 lensmeter to test lens data and convey the date to YPA 2100 5 Click key 6 Choose No00001...
Страница 80: ...75 7 Click key to complete data import...