22
Div&Con Test: on, off; factory default: on
Set if test the eyes’ collecting capability.
Printing mode: all, except for AR; U.S.F, U, LM, AR, SUBJ and FINAL; factory default: all
Set data content to be printed
All: all data
Except for AR: all data except for AR;
U.S.F: uncased eye, subjective and final data.
LM: lens meter data and vision functional test data;
AR: auto refractometer data and vision functional test data;
SUBJ: subjective and vision functional test data;
FINAL: final data and vision functional test data.
Date Format: M/D/Y, D/M/Y, Y/M/D; factory default: M/D/Y
Set printing data format.
Clear data after print: on, off; factory default: off
Set if delete measuring data after printing.
Clear data: on, off; factory default: off
Set if prompt “Confirm delete” message when pressing “delete”.
Set it “Yes” to avoid mistaken deletion of data when the “Delete” key is accidentally pressed.
CYL from LM: on, off; factory default: off
Import data from lens meter, set if automatically input prism data.
Fog:Balance :on, off; factory default: open
Set if automatic scieropia during vision test to keep the patient’s vision to 1.0 from 0.7.
AR data: on, off; factory default: off
Set if automatically receive computer refractor data.
LM data: on, off; factory default: off
Set if automatically receive lens meter data.
Close link: on, off; factory default: off
Avoid shifting assistant lens or mode through a corresponding sighting mark during optometric test.
Operation time: on, off; factory default: on
Содержание YPA-2100
Страница 1: ...YPA 2100 DIGITAL REFRACTOR User Manual Version 1 0 Date 20180521...
Страница 5: ...13 Electrical Schematic Diagram 82 14 Electromagnetic Compatibility 83...
Страница 24: ...19 Press to enlarge PD press to reduce PD 3 Press to exit PD input interface...
Страница 31: ...26 key...
Страница 42: ...37 4 Click key shift to 0 5XC 5 Click the key again shift to automatic mode Slit Prism...
Страница 55: ...50 34 Press to finish the test...
Страница 72: ...67 3 Program 1 A B C D are recommended vision test steps users can also add or delete as needed...
Страница 74: ...69 Add Key Click it to add one step automatically 4 Click key to enter program customizing interface...
Страница 78: ...73 3 Click key to popup input box...
Страница 79: ...74 4 Use CCQ 800 lensmeter to test lens data and convey the date to YPA 2100 5 Click key 6 Choose No00001...
Страница 80: ...75 7 Click key to complete data import...