17
4. Operating procedure
Fixed temperature operation
FIXED TEMP (Fixed Temperature) mode is to keep running at target temperature. It will keep running
until operation stops.
SV
: Set Value (Target Temperature), t
:
Time
Set Fixed Temperature mode.
1 Turn on the controller.
Turn on Earth Leakage Breaker(ELB) on (
|
) the left
side wall of this Equipment. (
Idle State
)
Press and hold key to turn on the controller power.
(
Standby State
)
Indicate read temperature in Furnace on Top Screen
and indicate target temperature on Bottom Screen.
The fan motor will start.
2 Set to the fixed temperature mode.
Press
key to turn FIXED TEMP (Fixed
Temperature mode) lamp on.
※
Fixed Temperature mode would be selected at first
time operation. After that, most latest operated
mode is selected.
3 Set a temperature.
①
Press
key. Then flash changeable digits on
Bottom Screen.
②
Shift to flashing digit with
key and then
change to desired digit with
keys.
Operating Temperature Range
:
0
~
1200
℃
(
FP102/302/312/412
)
③
Press
key when target temperature setting
has completed.
Press
key once or twice to cancel its setting.
SV
t
▲
Stop operation (manually)
▲
Start operation (manually)