6
2.
Before using this unit
Requirements for Installation
Warning
4. Do not modify
5. Installation on horizontal surface
Modification of this unit is strictly
prohibited. It may cause a failure.
Set this unit to the flattest place. Setting this
unit on rough or slope place could cause the
vibration
or
noise,
or
cause
the
unpredictable trouble or malfunction.
SINCE 1889
DP83C
Vacuum Drying Oven
改造
SINC
E 188
9
DP83
C
Vacu
um D
rying
Ove
n
6. Do not make an overload
7. Do not set samples in close formation
The withstand load of shelf is 15kg
(uniform load). Set the samples apart each
other.
The temperature in furnace cannot be
controlled if too much samples are set there.
Make sure to use the shelf and set samples
apart each other so as to make the free
space of 30% or more to the furnace to
acquire accuracy of temperature.
Make the free space of 30% or more
Samples
15kg
Shelf
Do not modify