19
4.
Operation Method
Quick Auto Stop Operation
Quick auto stop
operation procedure
This operation is used to specify the period up to automatic stop, i.e.,
sets the auto stop timer during operation.
1. Set the time up to stop during fixed temperature operation
Check that the FIXED TEMP lamp lights on and that the unit is
under operation.
Press the TIMER key.
The measurement temperature display screen displays the
character "tim", which indicates the timer setting. The setting
temperature display screen displays the current setting time with
blinking.
Select the time by pressing the "▼▲".
Timer function:
The maximum setting time is "999hours and 50 minutes ".
The time can be set in increments of a minute under 99 hours and
59 minutes.
It can be set in increment of ten minutes over 100 hours.
The "▼▲"can change the setting time quickly when it is pressed
continuously. Press them discontinuously when fine adjustment is
needed.
2. Start timer operation
Press the START/STOP key for 1 second after deciding the time.
Timer operation starts with the FIXED TEMP and AUTO STOP
lamps lighting on.
The timer is activated at the point when the START/STOP key is
pressed.
3. Stop/terminate timer operation
The operation stops automatically at setting time.
Buzzer continues to sound for about 5 seconds at operation stop.
The setting temperature screen displays the character "End", which
indicates termination of operation, with the FIXED TEMP and AUTO
STOP lamps lighting on. Press the START/STOP key to terminate
the timer operation mode. The screen returns to the initial setting
screen.