Handling Precautions
Caution
7. Setting of sample
Since the withstand load of the attached shelf plate is about 15kg per one plate, do not set
heavier sample than 15kg. When setting several sample, set them as dispersed as
possible. Too much sample setting could cause the improper control of the temperature. For
keeping the proper temperature, keep more than 30% space against whole size of the shelf
plate, and set the sample.
8.
Do not put sample on the chamber surface
If sample is put directly on the chamber surface, device performance will be disturbed.
Furthermore, internal temperature will abnormally rise and it will cause trouble. Never put
sample on the chamber surface. Fix the shelf on the bracket then set sample on it.
9. Recovering after power failure
When power is supplied after a power failure, the device automatically starts operation again
with the same state as just before the power failure. It is danger that the device starts
unattached operation after a power failure. We recommend for you to turn off the switch of
this unit if a power failure occurs during operation.
10.
Double stacking
Use the specified fittings included in the optional accessory for double stacking.
Do not make direct double stacking.
11.
After installation
It may cause injure to a person if this unit falls down or moves by earthquake or impact. To
prevent, take measures that the unit cannot fall down.