20
4.
Operation Method
Quick Auto Stop Operation
Change the set temp.
and set time, confirm
the set value
If need to change the set temp. during quick auto stop operation, press
FIXED TEMP key to enter the setting mode.
If need to change the set time during quick auto stop operation, press
TIMER key to enter the setting mode.
①
After changing the time, press TIMER.
The remaining time on the timer is the time subtracting the elapsed
time.
For example: set timer as 1hr30min, conduct quick auto stop
operation, after 20min, change the set time to be 2hr, set it by
TIMER, the remaining time is 1hr40min.
②
After changing the time, press START/STOP for 1sec.
The quick auto stop operation proceeds again as per the changed
time.
For example: set timer as 1hr30min, conduct quick auto stop
operation, after 20min, change the set time to be 2hr, press
START/STOP to set, the remaining time is 2hr.
Press
▼
key; switch over to display the set temp., operation mode
and remaining time of set temp. displayer.