33
4.
Operation Method
Other Functions (Calibration)
Use
calibration
offset function
Calibration offset is a function which corrects the difference between
the temperature in furnace and that of controller (sensor temperature)
if arises. The function parallel corrects the difference either to the plus
or minus side within the whole temperature range of unit. The
function can be set or cancelled by the SUBMENU key.
①
Start operation with the target setting temperature. Check the
temperature in furnace (temperature of sample) with a
thermograph after it is stabilized.
②
Check the difference between the setting temperature and that in
furnace (temperature of sample).
③
Press the SUBMENU key. Select the character "cAL", which
indicates the calibration offset, using the "▲▼", and then press
the ENTER key.
④
Input the difference using the "▲▼" and then press the ENTER
key. This completes the setting.
The setting range of offset correction temperature is +99
℃
to plus
side and -99
℃
to minus side respectively.
When it is set to the minus side, the temperature on the
measurement temperature display screen falls by the setting
temperature, while the temperature on furnace rises.
When it is set to the minus side, the temperature on the
measurement temperature display screen rises by the setting
temperature, while the temperature on furnace falls.
The unit has two-point correction function, which performs offset
between low-temperature zone and high-temperature zone.
Please consult our local branch office when carrying out validation
of temperature controller.