38
6.
Maintenance Method
Daily Inspection and Maintenance
Warning
●
Be sure to pull out the power cord except under special circumstances before trying to do
inspection and maintenance works.
●
Start these works after the device has returned to the normal temperature.
●
Never try to disassemble the unit.
Caution
Wipe off the dirt on resin parts or operation panel with a tightly wrung soft cloth. Never clean the
unit with gasoline, banana oil or detergent, otherwise it may result in deformation, deterioration or
discolor.
Every month
●Please inspect the ELB performance regularly.
Power on to test
Firstly, turn ELB on
Then, use a tip (like pencil tip) to press the test
switch of ELB. If ELB breaks off, it’s normal.
测试开关
Test switch