Operation Method
Program Operation
DPC type
temp. rise
/ fall time
The DPC type temp. rise / fall time is as below:
This value is the time needed to reach each temperature (eg. DP83C
needs about 15min rising temperature from 100
℃
to 150
℃
).
However, the constant temperature time reaching each set
temperature needs to be counted separately, so make sure to set
proper time to carry on test run.
Condition: RT
23
℃
humidity 65RH±20%
unloaded
(unit: min)
DP83C
Rise time
Fall time
200
℃
45
60
150
℃
35
180
100
℃
25
450
50
℃
15
-
Program creation
The program pattern below is explained as an example.
1.
Program pattern example
Temp.
200
℃
100
℃
0
℃
Step
1
2
3
4
5
6
7
← Repeat function →
The number of steps is not counted
8
9
Temp.(
℃
)
100 100 150 150 200
200
50
50
25
Time(min.)
25
30
15
90
25
180
150
180
100
2.
Turn on the power
Turn on the power switch of the unit.
The display on the controller lights on.
The initial screen is displayed for about 4 seconds, then the
measurement temperature (temperature in furnace) is displayed.