2

Ⅱ. SPECIFICATIONS OF MANOSTAR SWITCH MS65 AND MS65F

Types  :

Pressure ranges and characteristics  :

Pressure

range codes

Pressure ranges

Accuracy of 

switching point

(at 20℃ )

Max. switching 

difference pressures

Materials of 

pressure receiving

element

Pa

kPa

60D

 

10 ~  60

  (0.01 ~   0.06)   ±   2.0 

Pa

 

15.0 

Pa

Polyester fi lm

120D

 

20 ~ 120

  (0.02 ~   0.12)   ±   4.0 

Pa

 

20.0 

Pa

300D

 

50 ~ 300

  (0.05 ~   0.3 )   ±   8.0 

Pa

 

40.0 

Pa

600D

 

100 ~ 600

  (0.1  ~   0.6 )   ± 15 

Pa

 

60 

Pa

1.2E

   0.2  ~   1.2

  ±   0.030  kPa

 

  0.120  kPa

3E

   0.5  ~   3

  ±   0.080  kPa

 

  0.400  kPa

6E

   1 

~   6

  ±   0.15  kPa

 

  0.60  kPa

Silicone rubber

12E

   2 

~ 12

  ±   0.30  kPa

 

  1.20  kPa

30E

   5 

~ 30

  ±   0.75  kPa

 

  4.00  kPa

(   ) : Out of standard

㹋㹑㸴㸳 㹄 㹆 㹔㹊 㸱㸮㸮 㹂 㸫㹉

㸫㹉

䠬䠽

䡇䠬䠽


㹋㹊

㹔㹊

㻲㼛㼞㻌㼓㼑㼚㼑㼞㼍㼘㻌㼑㼘㼑㼏㼠㼞㼕㼏㼍㼘㻌㼘㼛㼍㼐
㻲㼛㼞㻌㼘㼛㼣㻌㼑㼘㼑㼏㼠㼞㼕㼏㼍㼘㻌㼘㼛㼍㼐

㼂㼀㻌㼟㼠㼞㼍㼕㼓㼔㼠㻌㼠㼥㼜㼑㻌

㼂䠨㻌㼞㼛㼠㼍㼞㼥㻌㼘㼛㼛㼟㼑㻌㼠㼥㼜㼑

㻹㼀㻌㼠㼟㼠㼞㼍㼕㼓㼔㼠㻌㼠㼥㼜㼑

㻹䠨㻌㼞㼛㼠㼍㼞㼥㻌㼘㼛㼛㼟㼑㻌㼠㼥㼜㼑

㻲㼛㼞㻌㼡㼜㼜㼑㼞㻌㼘㼕㼙㼕㼠㻌㼟㼑㼠㼠㼕㼚㼓
㻲㼛㼞㻌㼘㼛㼣㼑㼞㻌㼘㼕㼙㼕㼠㻌㼟㼑㼠㼠㼕㼚㼓

㻲㼛㼞㻌㼢㼕㼚㼥㼘㻌㼛㼞
㼜㼘㼍㼟㼠㼕㼏㻌㼠㼡㼎㼑

㻲㼛㼞㻌㼙㼑㼠㼍㼘
㼠㼡㼎㼑

㹀㹪㹟㹬㹩

㻹㼍㼤㻚㻌㼢㼍㼘㼡㼑㻌㼛㼒㻌㼟㼏㼍㼘㼑

㻼㼞㼑㼟㼟㼡㼞㼑㻌㼞㼍㼚㼓㼑㻌㼏㼛㼐㼑

䠄㼑㼤㼍㼙㼜㼘㼑䠅

㻯㼛㼚㼠㼍㼏㼠

㼁㼚㼕㼠

㻼㼞㼑㼟㼟㼡㼞㼑㻌㼞㼍㼚㼓㼑

㻼㼕㼜㼕㼚㼓㻌䠿㼛㼚㼚㼑㼏㼠㼛㼞

㻿㼑㼠㼠㼕㼚㼓㻌㼛㼒㻌㼟㼏㼍㼘㼑

㼀㼥㼜㼑


㼃㼕㼠㼔㻌㼍㼕㼞㻌㼒㼕㼘㼠㼑㼞

㹀㹪㹟㹬㹩

Содержание MS65F

Страница 1: ...I N S T R U C T I O N M A N U A L 653 0031 1 Nishi shiriike cho Nagata ku Kobe Japan TEL 078 631 6000 FAX 078 631 6020 READ THE INSTRUCTION MANUAL RoHS Compliant...

Страница 2: ...Caution of piping 6 7 Installation position 6 SETTING OF PRESSURE 1 Switch contact con guration 7 2 Setting of upper limit lower limit 7 3 Reset time 8 ABOUT BUILT IN MICRO SWITCH 1 Material of conta...

Страница 3: ...instrument The instrument is measurable for air and non corrosive gas only The machine is exclusive use of dry air 90 RH or less Using measuring the water or oil it may be damaged and causes the accid...

Страница 4: ...ce pressures Materials of pressure receiving element Pa kPa 60D 10 60 0 01 0 06 2 0 Pa 15 0 Pa Polyester lm 120D 20 120 0 02 0 12 4 0 Pa 20 0 Pa 300D 50 300 0 05 0 3 8 0 Pa 40 0 Pa 600D 100 600 0 1 0...

Страница 5: ...de 10 10 to 50 Hz Acceleration 39 m s2 each two hours on triaxial directions 100 m s2 each six times on triaxial direction Between each terminal and case 20 M or more 500 V DC megger Between each term...

Страница 6: ...trical load type The general electrical load type is decreased contact reliability refer to p 8 The withstanding pressure of the pressure receiving element means The max pressure under which the diaph...

Страница 7: ...nt of negative pressure Connect the tube to the low pressure side piping connector blue or L The high pressure port red or H should be opened to atmosphere but do not remove the piping connector c Mea...

Страница 8: ...it once in a while After the cleaning of the tank check that the air tightness is fully kept c Measurement of high temperature gases In the pressure measurement of high temperature gas use the pressur...

Страница 9: ...pressure the other is called lower limit setting type type L which is adjusted on reset pressure Be careful about the difference between the upper and lower limit setting switched and also understand...

Страница 10: ...contact for low electrical load cross point contact Magni ed drowing of contact for general load small rivet contact point 2 Contact material and applicable mecanism for low voltage and low current c...

Страница 11: ...ive force of inductive load by diode and avoid to apply high voltage to inbetween the contacts This method makes the return time of inductive load slower Choose the diode rated current is more than lo...

Страница 12: ...acid and it may also cause electrical contact failure or malfunction of switches b Silicone gas If the gas to be measured contains silicon produced from silicon contained in oil grease filling agent...

Страница 13: ...ain to x Replacement of air lter In case of using model with air lter MS65F depending on amount of work conduct disassembly and clean out once in 2 years periodically and replace lters and packings wi...

Страница 14: ...rument life For vinyl or plastic tube For metal tube Product code FG VT6 S Product code FG MT6 S T joint For vinyl or plastic tube Product code TVT6 4 6 For branching different I D 6 and 4 Product cod...

Страница 15: ...og speci cations or manual Handling or usage of the product other than as described in our product catalog speci cations or manual 2 Breakdown due to a reason other than a fault within our product 3 A...

Отзывы: