Oxi 3310
Operation
27
ba75804e08
10/2017
4.4
Dissolved oxygen
4.4.1
General information
You can measure the following parameters:
D.O. concentration
D.O. saturation index ("D.O. saturation")
D.O. partial pressure
D. O. measurements with the Oxi 3310 can be carried out with the
CellOx 325
and
DurOx 325
D.O. sensors.
The measuring instrument automatically recognizes the type of the
connected D.O. sensor.
Attention
When a grounded PC is connected, measurements cannot be per-
formed in grounded media as incorrect values would result.
The USB interface is not galvanically isolated.
Temperature
measurement
The D. O. sensors
CellOx 325
and
DurOx 325
have an integrated tem-
perature sensor.
Preparatory activities
Perform the following preparatory activities when you want to measure:
Note
Incorrect calibration of D.O. sensors will result in incorrect measured
values. Calibrate at regular intervals.
1
Connect the D.O. sensor to the meter.
The D.O. measuring screen is displayed.
2
Calibrate or check the meter with the sensor.
Содержание WTW Oxi 3310
Страница 1: ...OPERATING MANUAL ba75804e08 10 2017 Oxi 3310 DISSOLVED OXYGEN D O METER...
Страница 2: ...2 ba75804e08 10 2017 Oxi 3310 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 5: ...Oxi 3310 Contents 5 ba75804e08 10 2017...
Страница 10: ...Overview Oxi 3310 10 ba75804e08 10 2017...
Страница 14: ...Safety Oxi 3310 14 ba75804e08 10 2017...
Страница 48: ...Operation Oxi 3310 48 ba75804e08 10 2017...
Страница 51: ...Oxi 3310 Maintenance cleaning disposal 51 ba75804e08 10 2017...
Страница 53: ...Oxi 3310 What to do if 53 ba75804e08 10 2017...
Страница 56: ...Technical data Oxi 3310 56 ba75804e08 10 2017...
Страница 60: ...Lists Oxi 3310 60 ba75804e08 10 2017...
Страница 62: ...Firmware update Oxi 3310 62 ba75804e08 10 2017...
Страница 63: ......