Displaying calibration records
The calibration data can be displayed and then output to the interface.
Displaying the
calibration record
The calibration record of the last calibration is to be found under the
menu item,
Calibration
/
Calibration record
. To open it in the measured
value display, press the
<CAL
_
>
key.
4
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
5
Wait for the end of the AutoRead measurement or accept the
calibration value with
<ENTER>
.
The factor that was set last is displayed.
6
Using
<
> <
>
, set the correction factor to adjust the dis-
played concentration value to the nominal value (value of the
comparison measurement). Subsequently, accept the correc-
tion factor with
<ENTER>
.
The meter switches to the measured value display.
The status display
[Factor]
is active.
O
2
7.92
mg/l
25.1
°C
OxiComp
15.03.2014
08:00
O
2
7.92
mg/l
0.973
OxiComp
15.03.2014
08:00
Содержание WTW Oxi 3310
Страница 1: ...OPERATING MANUAL ba75804e08 10 2017 Oxi 3310 DISSOLVED OXYGEN D O METER...
Страница 2: ...2 ba75804e08 10 2017 Oxi 3310 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 5: ...Oxi 3310 Contents 5 ba75804e08 10 2017...
Страница 10: ...Overview Oxi 3310 10 ba75804e08 10 2017...
Страница 14: ...Safety Oxi 3310 14 ba75804e08 10 2017...
Страница 48: ...Operation Oxi 3310 48 ba75804e08 10 2017...
Страница 51: ...Oxi 3310 Maintenance cleaning disposal 51 ba75804e08 10 2017...
Страница 53: ...Oxi 3310 What to do if 53 ba75804e08 10 2017...
Страница 56: ...Technical data Oxi 3310 56 ba75804e08 10 2017...
Страница 60: ...Lists Oxi 3310 60 ba75804e08 10 2017...
Страница 62: ...Firmware update Oxi 3310 62 ba75804e08 10 2017...
Страница 63: ......