Oxi 3310
Operation
33
ba75804e08
10/2017
the calibration. The evaluation appears on the display and in the cali-
bration record.
Calibration in
water vapor
saturated air
(air calibration vessel)
For this calibration procedure, the
Comparison meas.
setting must be
set to
Off
in the
Calibration
menu.
Proceed as follows to calibrate the instrument:
Note
The sponge in the air calibration vessel must be moist (not wet). Leave
the sensor in the air calibration vessel for a time long enough to adjust.
Display
Calibration record
Relative slope
+++
S = 0.8 ... 1.25
++
S = 0.7 ... 0.8
+
S = 0.6 ... 0.7
Error
Error
S < 0.6 or S > 1.25
Eliminate the error according to
chapter 6 W
1
Connect the D.O. sensor to the meter.
2
Put the D.O. sensor into the air calibration vessel.
3
Start the calibration with
<CAL>
.
The last calibration data (relative slope) is displayed.
Содержание WTW Oxi 3310
Страница 1: ...OPERATING MANUAL ba75804e08 10 2017 Oxi 3310 DISSOLVED OXYGEN D O METER...
Страница 2: ...2 ba75804e08 10 2017 Oxi 3310 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 5: ...Oxi 3310 Contents 5 ba75804e08 10 2017...
Страница 10: ...Overview Oxi 3310 10 ba75804e08 10 2017...
Страница 14: ...Safety Oxi 3310 14 ba75804e08 10 2017...
Страница 48: ...Operation Oxi 3310 48 ba75804e08 10 2017...
Страница 51: ...Oxi 3310 Maintenance cleaning disposal 51 ba75804e08 10 2017...
Страница 53: ...Oxi 3310 What to do if 53 ba75804e08 10 2017...
Страница 56: ...Technical data Oxi 3310 56 ba75804e08 10 2017...
Страница 60: ...Lists Oxi 3310 60 ba75804e08 10 2017...
Страница 62: ...Firmware update Oxi 3310 62 ba75804e08 10 2017...
Страница 63: ......