What to do if...
HandyLab 100
52
ba77062d02
12/2017
Obviously incorrect
measured values
Meter does not react to
keystroke
You want to know which
software
version is in the meter
Cause
Remedy
pH electrode:
– pH electrode unsuitable
– Use suitable electrode
– Temperature difference
between buffer and test sam-
ple too great
– Adjust temperature of buf-
fer or sample solutions
– Measurement procedure not
suitable
– Follow special procedure
Cause
Remedy
– Operating condition unde-
fined or EMC load unallowed
– Processor reset:
Press the
<ENTER>
and
<On/Off>
key simultane-
ously
Cause
Remedy
– E.g., a question by the ser-
vice department
– Switch on the meter.
Open the menu,
<F1
_
>
/
[Menu] /
Storage & config
/
SystemService informa-
tion
. The instrument data
are displayed.