Operation
HandyLab 100
36
ba77062d02
12/2017
Note
For
single-point calibration
, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the electrode.
Continuing with two-
point calibration
Continuing with three-
to five-point calibration
8
If necessary, finish the calibration procedure as a single-point
calibration with
<M>
.
The calibration record is displayed.
9
Thoroughly rinse the electrode with deionized water.
10
Immerse the electrode in buffer solution 2.
11
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with
<
><
>
.
12
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
13
Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with
<ENTER>
.
The calibration display for the next buffer appears (voltage dis-
play).
14
If necessary, finish the calibration procedure as a two-point cal-
ibration with
<M>
.
The calibration record is displayed.
pH
0.0
24.8 °C
AutoCal TEC
7.000
15.03.2014
08:00
AR
Buffer
mV
15
Thoroughly rinse the electrode with deionized water.
16
Immerse the electrode in the next buffer solution.