What to do if...
Error message
OFL, UFL
Error message,
Error
Cause
Remedy
pH electrode:
– Measured value outside the
measuring range
– Use suitable electrode
– Air bubble in front of the junc-
tion
– Remove air bubble
– Air in the junction
– Extract air or moisten junc-
tion
– Cable broken
– Replace the electrode
– Gel electrolyte dried out
– Replace the electrode
Cause
Remedy
pH electrode:
– The values determined for
zero point and slope of the
electrode are outside the
allowed limits.
– Recalibrate
– Junction contaminated
– Clean the junction
– Electrode broken
– Replace the electrode
Buffer solutions:
– Incorrect buffer solutions
– Change calibration proce-
dure
– Buffer solutions too old
– Use only once.
Note the shelf life
– Buffer solutions depleted
– Change solutions