Operation
HandyLab 100
32
ba77062d02
12/2017
* Brand names or trade names are trademarks of their respective owners
protected by law.
Note
The buffers are selected in the menu, pH /
<F1>
/[Menu] /
Calibration
/
Buffer
(see page 28).
Calibration points
Calibration can be performed using one to five buffer solutions in any
order (single-point to five-point calibration). The meter determines the
following values and calculates the calibration line as follows:
20
Reagecon TEC
*
2.000
4.010
7.000
10.000
25 °C
21
Reagecon 20
*
2.000
4.000
7.000
10.000
13.000
20 °C
22
Reagecon 25
*
2.000
4.000
7.000
10.000
13.000
25 °C
23
Chemsolute *
4.000
7.000
10.000
20 °C
24
USABlueBook
*
4.000
7.000
10.000
20 °C
25
YSI
*
4.000
7.000
10.000
25 °C
No.
Buffer set *
pH values
at
Determined values
Displayed calibration data
1-point
Asy
Zero point =
Asy
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
2-point
Asy
Slp.
Zero point =
Asy
Slope =
Slp.