HandyLab 100
Operation
39
ba77062d02
12/2017
Note
For
single-point calibration
, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the electrode.
Continuing with two-
point calibration
8
Set the nominal buffer value for the measured temperature with
<
><
>
.
9
Accept the calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage dis-
play).
10
If necessary, finish the calibration procedure as a single-point
calibration with
<M>
.
The calibration record is displayed.
pH
7.000
24.8 °C
VariCal
1
15.03.2014
08:00
Buffer
11
Thoroughly rinse the electrode with deionized water.
12
Immerse the electrode in buffer solution 2.
13
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with
<
><
>
.
14
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
15
Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with
<ENTER>
.
The calibration display for the next buffer appears.
Содержание SI Analytics HandyLab 100
Страница 1: ...OPERATING MANUAL ba77062d02 12 2017 HandyLab 100 pH METER HandyLab 100...
Страница 58: ...Lists HandyLab 100 58 ba77062d02 12 2017...
Страница 59: ......