Operation
HandyLab 100
40
ba77062d02
12/2017
Continuing with three-
to five-point calibration
16
Set the nominal buffer value for the measured temperature with
<
><
>
.
17
Accept the calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage dis-
play).
18
Finish the calibration procedure as a two-point calibration with
<M>
.
The calibration record is displayed.
pH
4.035
24.8 °C
VariCal
2
15.03.2014
08:00
Buffer
19
Thoroughly rinse the electrode with deionized water.
20
Immerse the electrode in the next buffer solution.
21
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with
<
><
>
.
22
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
23
Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with
<ENTER>
.
The calibration display for the next buffer appears.