Operation
HandyLab 100
38
ba77062d02
12/2017
1
In the measured value display, select the measured parameter
pH or mV with
<M>
.
2
Start the calibration with
<CAL>
.
The calibration display appears.
3
Thoroughly rinse the electrode with deionized water.
4
Immerse the electrode in buffer solution 1.
5
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with
<
><
>
.
6
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
7
Wait for the end of the measurement with stability control or
accept the calibration value with
<ENTER>
.
The calibration display for the next buffer appears.
pH
0.0
24.8 °C
VariCal
1
15.03.2014
08:00
Buffer
mV
pH
0.0
24.8 °C
VariCal
7.000
15.03.2014
08:00
Buffer
mV
AR