Operation
Power Supply Operation
Release 2.1
57
Power On
Output State
(XFR/XHR only) You may also change the output state whether the output is enabled
or disabled at power on.
To change the power on at output state:
OUTP:PON:STAT [ON|OFF|1|0]
Whether the unit powers on with output enabled is a function of both the state at
power off and this setting.
Reset
Resetting the unit puts certain features to a known state. These states are listed in
Table 3.2
.
To reset the unit:
*RST or SYST:RES
Table 3.2
Features Affected by Reset (*RST) Command
Feature
Reset State (*RST)
Voltage setpoint
0.0V
Current setpoint
0.0A
Low voltage setpoint limit
0.0V
High voltage setpoint limit
101% of voltage rating
Low current setpoint limit
0.0A
High current setpoint limit
101% of current rating
Over voltage protection
Disabled (0.0V)
Under voltage protection
Disabled (0.0V) and not shutdown when tripped
Over current protection
Disabled (0.0A) and not shutdown when tripped
Under current protection
Disabled (0.0A) and not shutdown when tripped
Fold shutdown protection
None and delay 0.5s
AC off shutdown recovery
Auto recover
OTP shutdown recovery
Latched
Triggered voltage setpoint
Disabled (Default)
Triggered current setpoint
Disabled (Default)
Autosequence mode
Exit
Calibration mode
False
Output
Disabled
Содержание GPIB-M-HPD
Страница 2: ......
Страница 18: ...List of Figures xvi Operating Manual for Multichannel Functionality GPIB M...
Страница 26: ...Features and Specifications Specifications 24 Operating Manual for Multichannel Functionality GPIB M...
Страница 94: ...Current Sharing Operation 92 Operating Manual for Multichannel Functionality GPIB M...
Страница 98: ...GPIB Performance Specifications 96 Operating Manual for Multichannel Functionality GPIB M...
Страница 118: ...SCPI Command Reference Expressions 116 Operating Manual for Multichannel Functionality GPIB M...
Страница 132: ...Calibration Restore Factory Calibration 130 Operating Manual for Multichannel Functionality GPIB M...
Страница 133: ......