Operation
Remote Control Modes
46
Operating Manual for Multichannel Functionality (GPIB-M)
Remote Control Modes
The power supply must be in remote mode to execute commands. The remote mode
is indicated by the front panel remote status LED. A power supply will respond to
queries when in local mode, but will generate an error when a command is received.
GPIB equipped supplies will automatically transition from local to remote mode when
any command data is received. CANbus-only supplies may have to be put into remote
mode with a command before sending any other command. See
“Change
Remote/Local Mode” on page 48
.
The factory default power-on settings are given in
Table 3.1, on page 46
.
Table 3.1
Power Supply Factory Defaults
You can change the default Remote/Local state, see
“Power On Remote State” on page
49
.
If safety is an issue because the front panel control knobs may be set to hazardous
voltages at power on, contact the factory to order locking bushings for the knobs.
See your power supply’s operating manual for details on using the power supply in
local mode.
Power Supply
Default Remote State
Default Output State
(Shutdown)
XFR or XHR
Remote
OFF
XPD
Local
ON
XT or HPD
Local
ON
Содержание GPIB-M-HPD
Страница 2: ......
Страница 18: ...List of Figures xvi Operating Manual for Multichannel Functionality GPIB M...
Страница 26: ...Features and Specifications Specifications 24 Operating Manual for Multichannel Functionality GPIB M...
Страница 94: ...Current Sharing Operation 92 Operating Manual for Multichannel Functionality GPIB M...
Страница 98: ...GPIB Performance Specifications 96 Operating Manual for Multichannel Functionality GPIB M...
Страница 118: ...SCPI Command Reference Expressions 116 Operating Manual for Multichannel Functionality GPIB M...
Страница 132: ...Calibration Restore Factory Calibration 130 Operating Manual for Multichannel Functionality GPIB M...
Страница 133: ......