Operation
Auto Sequencing
Release 2.1
65
Set V, I, and P
Limits
The voltage, current and power setpoints can be limited to less than the supply rating
range to match the tolerance of connected equipment or any other criteria you may
have.
Once the limits have been changed from the supply’s default rated output, settings
outside this range are no longer accepted.
To set the voltage limits:
SOUR:VOLT:LIM:HIGH <voltage>
SOUR:VOLT:LIM:LOW <voltage>
To set the current limits:
SOUR:CURR:LIM:HIGH <current>
SOUR:CURR:LIM:LOW <current>
If the high end of the range was set to 5 volts, the command, SOUR:VOLT 10 would
return an error.
Triggered
Setpoints
To set a triggered setpoint:
SOUR:VOLT:TRIG <voltage>
SOUR:CURR:TRIG <current>
See
“Triggering Commands” on page 66
for more information.
Note
Setpoint limits do not apply to the triggered outputs and auto sequence outputs.