Operation
Power Supply Operation
Release 2.1
55
User Settings
(Save and
Recall)
If you have a frequent or constant need for a specific voltage and current output, you
can save these setpoints in the power supply’s memory as a user setting. Once a
setting is stored, it remains in the power supply’s memory after the unit is powered
off.
Ten user setting memory locations are available, and each saves the following
parameters:
•
Voltage setpoint
•
Current setpoint
•
Over-voltage protection setpoint (OVP)
•
Under-voltage protection setpoint and shutdown configuration (UVP)
•
Over-current protection setpoint and shutdown configuration (OCP)
•
Under-current protection setpoint and shutdown configuration (UCP)
•
Foldback protection mode and delay settings
•
AC Off auto recovery configuration (ACO)
•
OTP auto-recovery configuration
•
Aux line configuration
•
Voltage and current soft limits
•
Triggered setpoints and trigger source
To store (save) settings:
*SAV <user-setting>
or the SCPI equivalent for multichannel addressed units
SYST<channel>:SAVE <user-setting>
where <user-setting> is an integer from 1 to 10.
Example:
*SAV 1
will save all the present settings to user setting location 1.
To recall settings:
*RCL <user-setting>
To save default settings:
*SDS <user-setting>
Note
All parameters are saved and used when the user setting is recalled. Therefore, you
should set parameters that you do not care about to the factory defaults.
Содержание GPIB-M-HPD
Страница 2: ......
Страница 18: ...List of Figures xvi Operating Manual for Multichannel Functionality GPIB M...
Страница 26: ...Features and Specifications Specifications 24 Operating Manual for Multichannel Functionality GPIB M...
Страница 94: ...Current Sharing Operation 92 Operating Manual for Multichannel Functionality GPIB M...
Страница 98: ...GPIB Performance Specifications 96 Operating Manual for Multichannel Functionality GPIB M...
Страница 118: ...SCPI Command Reference Expressions 116 Operating Manual for Multichannel Functionality GPIB M...
Страница 132: ...Calibration Restore Factory Calibration 130 Operating Manual for Multichannel Functionality GPIB M...
Страница 133: ......