Electrical Interface Details
Power/Comm/Control Cable
Item Pin Name
Description Active
State
Comments
1
J1 -1
Power
+24 Vdc
-
2.45 Amp max.
1.40 Amp Typ.
2
J1 -2
Negative
Power Return
-
Negative side of 24Vdc power
supply
3
J1 -3
Tx out
RS232 Transmit Out
- RS232-TX
4
J1 -4
RV in
RS232 Receive In
-
RS232-RV
5
J1 -5
DSR
Discrete Select
Request
High
See Note 1
Input +24V max, +12V min.
rating, 100mA max.
6
J1 -6
10's
Ten's Digit
High (50 msec.
pulse)
DSR must be active to work
Input +24V max, +12V min
rating, 100mA max.
7
J1 -7
1's
One's Digit
High (50 msec.
pulse)
DSR must be active to work
Input +24V max, +12V min
rating, 100mA max.
8
J1 -8
Read
Take a
Measurement
High (50 msec.
pulse)
DSR must be active to work
Input +24V max, +12V min
rating, 100mA max.
Cable
Shield
-
Shield
-
-
Connect to the panel or fixture
ground
Discrete Interface Cable
Item Pin
Name
Default Option
(see Note 2)
Description Active
State
Comments
1
J2 -1
Pass
Pass
Pass output
See Note 3
2
J2 -2
Fail
Fail
Fail output
See Note 3
3 J2
-3
Output
Common
Output
Common
Output Common
(see Note 3)
See Note 3
4
J2 -4
ACTION
CH-E 16
ACTION and/or
Channel-E
See Note 3
5
J2 -5
CH-D 8
RS485TX-A
Channel-D or
RS485-TX-A
See Note 3
6
J2 -6
CH-C 4
RS485RX-A
Channel-C or
RS485-RX-A
See Note 3
7
J2 -7
CH-B 2
RS485RX-B
Channel-B or
RS485-RX-B
See Note 3
8
J2 -8
CH-A 1
RS485TX-B
Channel-A or
RS485-TX-B
See Note 3
Cable Shield
-
Shield
-
-
Connect to
the panel or
fixture
ground