Calibration Configuration
The configuration dialog is used to enter the white verification tolerance value and select the
mounting geometry. The other options are only edited if a replacement cal tube or reference
is used.
To access the Configuration dialog, click the
Configuration
button on the Calibration dialog.
Entering a White Verification Tolerance
The tolerance value entered is used to determine the verification difference tolerance. The
default value is set to 1.00.
1.
Click in the
White Verification Tol
field and edit the value.
2.
Click the
Load New Values
button to download to the instrument.
3.
Click the
Close
button to close the Configuration Calibration Values dialog.
Selecting the Mounting Geometry
This option is used to select the mounting geometry of the VeriColor Spectro.
1.
Select
0°/30°
or
30°/0°
option.
2.
Click the
Close
button to close the Configuration Calibration Values dialog.