OxiTop
®
Control OC 110
What to do if...
ba31116e05
07/2014
95
Check of the controller
Controller information
All single information on the state of the controller is listed. The check can
be undertaken at any time without affecting the measurements that are run-
ning.
Perform the command:
- All
. See
GLP/TOOLS
- Check
- Show
.
(The check is used in this case to test the IR interface.) All working measur-
ing heads must flash for 5 seconds.
Perform a measuring head check (see section 9.5.2 M
, page 70). (The check is used in this case to test the IR inter-
face.) Here, a working measuring head should supply its status data.
Keyboard, display and signal tone should show the required reaction.
Check the function of the clock (prerequisite: Clock is set).
Switch off the instrument and switch it on again. Time and date each appear
showing the current values.
Check RS232 interface (only required when using the WTW software,
ACHAT OC):
– Connect the controller to your PC by means of the interface cable
AK540/B.
– Switch on the controller and start the PC program, ACHAT OC. In doing
so, the PC program checks the RS232 interface.
The controller does not
react / is defective – the
data of the running mea-
surements should be
restored
Send the controller to WTW. Using a repaired or new controller (memory must
be empty!) perform the “Restore data” procedure (see section 9.6 M
Incorrect time display on
being switched on
Data backup battery was changed.
Summer/winter time change has
taken place.
Set the clock (see section 9.4.3
D
Note: The date and time must not be
changed in the controller between the
start and end of a sample so momen-
tary values can be correctly stored in
the
Pressure p
mode.
Cause
Remedy
Содержание OxiTop Control OC 110
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Contents OxiTop Control OC 110 6 ba31116e05 07 2014...
Страница 10: ...Overview OxiTop Control OC 110 10 ba31116e05 07 2014...
Страница 16: ...Commissioning OxiTop Control OC 110 16 ba31116e05 07 2014...
Страница 22: ...Operation OxiTop Control OC 110 22 ba31116e05 07 2014...
Страница 100: ...OxiTop Control OC 110 100 ba31116d05 07 2014...
Страница 101: ......