9.5.5
Pneumatic test
The pneumatic test tests the measurement precision of the measuring
head.
It says nothing about the long-term impermeability of the system.
To perform the pneumatic test, you require the test resource, OxiTop
®
PT (WTW order number 209 334).
The user interface of the controller guides you through the test:
CHECK
menu.
Use
<
><
>
to select the
-
Pneumatic test
menu item.
Set the plunger of the syringe on
the OxiTop
®
PT test resource to
5 scale parts.
Tightly screw the measuring
head to be tested onto the Oxi-
Top
®
PT test resource.
(Exceeding the time causes a
return to the
CHECK
menu).
(Exceeding the time causes a
return to the
CHECK
menu).
The controller display shows the
result of the pneumatic test.
Use
<RUN/ENTER>
to return to
the
CHECK
menu.
RUN/ENTER
RUN/ENTER
RUN/ENTER