9.4
Settings
With this function you can undertake or change settings. The table
below shows the settings in the delivery state.
GLP/TOOLS
menu.
Select the
- Settings
menu item with
<
><
>
GLP
Setting
Default setting
(bold)
and setting range
Operating mode
BOD Standar
d
BOD Routine
BOD Special
Pressure p
Operational life
5 days
(0.5 h to
99 days)
5 days
(0.5 h to
99 days)
5 days
(0.5 h to 99
days)
5 days
(0.5 h to 99
days)
Date
current
current
current
current
Time
current
current
current
current
GLP
On
Off
Off
(permanently
set)
On
Off
On
Off
Calibration interval
12 months
to 36 months
---
12 months
to 36 months
12 months
to 36 months
Clearing the memory
automatic
or manual
automatic
or manual
automatic
or manual
automatic
or manual
AutoTemp
On
Off
On
(permanently
set)
On
Off
Off
(permanently set)
Switch-off interval
5 minutes
to 15 minutes
5 minutes
(permanently set)
5 minutes
to 15 minutes
5 minutes
to 15 minutes
Temperature
20 °C
(permanently
set)
20 °C
(permanently set)
20 °C
(5 °C to 40 °C)
---
Height
500 m
(permanently
set)
500 m
(permanently set)
500 m
(-698 - 5572 m)
---
Air pressure
954 hPa
(permanently
set)
954 hPa
(permanently set)
954 hPa
(500 - 1100 hPa)
---
Limit pressure
---
---
---
150 hPa
(50 - 500 hPa)
Language
German
German
German
German
Содержание OxiTop Control OC 110
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Contents OxiTop Control OC 110 6 ba31116e05 07 2014...
Страница 10: ...Overview OxiTop Control OC 110 10 ba31116e05 07 2014...
Страница 16: ...Commissioning OxiTop Control OC 110 16 ba31116e05 07 2014...
Страница 22: ...Operation OxiTop Control OC 110 22 ba31116e05 07 2014...
Страница 100: ...OxiTop Control OC 110 100 ba31116d05 07 2014...
Страница 101: ......