The measurement principle
The respirometric measurement is a pressure measurement. If oxygen is
consumed in a closed vessel at a constant temperature, a negative pressure
develops. If a gas is released, an overpressure develops. The OxiTop
®
measuring head measures and stores this pressure for the whole duration of a
measurement once started.
The OxiTop
®
OC 110 controller collects the pressure values from the
measuring heads and processes them.
The formula shown below is the basis for all calculations for the BOD using the
values from the OxiTop
®
measuring head.
The interpretation of the pressure differences in the temporal course depends
on the measured material and its preparation and on the sample manipulation
(e. g. intermediate aerations) during the measuring period.
1 Measuring head OxiTop
®
-C or OxiTop
®
-C/B
®
®
®
®
®
®
1
2
0
m
l
l
t
m
2
O
p
T
T
V
V
V
T
R
O
M
BOD
M(O
2
)
Molecular weight (32000 mg/mol)
R
Gas constant (83.144 l
mbar/mol
K)
T
0
Reference temperature (273.15 K)
T
m
Measuring temperature
V
t
Bottle volume (nominal volume in ml)
V
l
Sample volume in ml
a
Bunsen absorption coefficient (0.03103)
p(O
2
)
Difference of the D. O. partial pressure (mbar)
Содержание OxiTop Control OC 110
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Contents OxiTop Control OC 110 6 ba31116e05 07 2014...
Страница 10: ...Overview OxiTop Control OC 110 10 ba31116e05 07 2014...
Страница 16: ...Commissioning OxiTop Control OC 110 16 ba31116e05 07 2014...
Страница 22: ...Operation OxiTop Control OC 110 22 ba31116e05 07 2014...
Страница 100: ...OxiTop Control OC 110 100 ba31116d05 07 2014...
Страница 101: ......