Operating mode: BOD Special
OxiTop
®
Control OC 110
42
ba31116e05
07/2014
In the
SELECT CALCULATION
menu you can select one parameter
out of bottle volume,
filling volume and measuring range to be calculated by the controller
based on the other parameters. Enter the respective other parameters
using
<
><
>
or confirm the presettings using
<RUN/ENTER>
.
When doing this the parameter to be calculated is changed simultane-
ously.
Example:
Change parameters to:
Filling volume of the sample to be calculated by the controller,
Bottle volume 900 ml, dilution 1 + 1, measuring range 2000 mg/l.
Altitude
500 m
-698 m to 5572 m (setting: see sec-
tion 9 GLP/TOOLS, page 57)
Parameters
Default setting Admissible setting range
With
<
><
>
select the
parameter the controller will
calculate.
The controller stores the set-
tings.
Preselected
in the delivery state:
-
Measuring RANGE
.
- Reset parameters
resets
the settings to the delivery
state.
After confirming, the display
shows the values last set up
for bottle volume, filling vol-
ume of the sample, dilution
(these can be changed), and
the measuring range that the
controller calculated from
these.
# marks the parameter calcu-
lated by the controller.
RUN/ENTER
RUN/ENTER
SELECT CALCULATION
menu:
Select the
- FILLING volume
menu item with
<
><
>
Содержание OxiTop Control OC 110
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...Contents OxiTop Control OC 110 6 ba31116e05 07 2014...
Страница 10: ...Overview OxiTop Control OC 110 10 ba31116e05 07 2014...
Страница 16: ...Commissioning OxiTop Control OC 110 16 ba31116e05 07 2014...
Страница 22: ...Operation OxiTop Control OC 110 22 ba31116e05 07 2014...
Страница 100: ...OxiTop Control OC 110 100 ba31116d05 07 2014...
Страница 101: ......