OxiTop
®
Control OC 110
Operating mode: Pressure p
ba31116e05
07/2014
51
For the menu items
- Show sample
,
- Erase sample
-
list
and
- Call
up data
refer to the corresponding paragraphs in section 5 O
7.6
Calling up momentary values
With this function you can call up and store the momentary pressure
value of a sample. The controller can store up to 10 momentary values
per sample. They appear as vertical lines on the curve in the evalua-
tion.
According to the sample selected, one of the two following
submenus appears.
Finished sample
Current sample
The header line contains:
Sample number
ID number (e.g. I001)
Type and run duration
of the selected sample.
RUN/ENTER
The date and time must not be changed in the controller between
the start and end of a sample so the momentary values can be cor-
rectly stored.
SAMPLE STATUS TYPE
(sam-
ple management):
Use
<
><
>
to select a sam-
ple.
A list with menu items for the
sample appears.
RUN/ENTER