Operation
Oxi 330i/340i
42
ba52306e04
03/2008
4.7
Reset
You can reset (initialize) the measurement parameters and the configuration
parameters separately from one another.
Measurement
parameters
The following measured parameters
(
O
2
InI)
are reset to the default condi-
tion:
Note
The calibration data gets lost when the measuring parameters are reset.
Recalibrate after performing a reset.
Configuration
parameters
The following configuration parameters (
InI
) are reset to the delivery status:
Resetting the
measuring
parameters
Measuring mode
D. O. concentration
AutoRange automatic measure-
ment range selection
On (
YES
)
Relative slope
1.00
Correction factor of the relative
slope
1.000
Salinity equivalent
0.0
Salt content correction
off
Baud rate
4800
Interval 1 (automatic save)
OFF
Interval 2 (for data transmission)
OFF
1
Press the
<RUN/ENTER>
key and hold it down.
2
Press the
<CAL>
key.
UpH
ARng
Y
Sal
Sal
/
K
%
TP
C
°
°
F
Store
cm
1/
1
S
M
m
%
mg/l
cm
S/
cm
mV/pH
mbar
Baud
4
Day.Month No.
Time
Auto
LoBat
Tref25
nLF
RCL
AR
Cal
Lin
Tref20
TDS
AutoCal DIN
AutoCal TEC
Ident
Year
o
n
I
n
I
O
Содержание Ox i340i
Страница 4: ...List of contents Oxi 330i 340i 4 ba52306e04 03 2008 6 What to do if 47 7 Technical data 51 8 Lists 55...
Страница 44: ...Operation Oxi 330i 340i 44 ba52306e04 03 2008...
Страница 50: ...What to do if Oxi 330i 340i 50 ba52306e04 03 2008...
Страница 60: ...Index Oxi 330i 340i 60 ba52306e04 03 2008...