Oxi 330i/340i
Operation
23
ba52306e04
03/2008
4.3.2
Calibrating via a comparison measurement
Proceed as follows to calibrate the instrument:
1
Connect the D. O. probe to the measuring instrument.
2
Press the
<CAL>
key repeatedly until
O
2
FAC
appears.
3
Immerse the D. O. probe in the calibrating solution.
4
Press the
<RUN/ENTER>
key.
The AutoRead measurement begins. If the measured value is stable,
the instrument displays the determined measured value and the cor-
rection factor of the relative slope (initial value = 1.000).
5
Remove the D. O. probe from the calibrating solution.
6
Perform a comparison measurement, e. g. Winkler titration, with the
calibrating solution (determine the nominal value). Leave the mea-
suring instrument switched on until the nominal value is determined.
7
Set the displayed value on the measuring instrument to the nominal
value with
<
▲
> <
▼
>
. This changes the related correction factor of
the relative slope.