Technical data
Oxi 330i/340i
52
ba52306e04
03/2008
Accuracy
(± 1 digit)
Correction
functions
Analog output
(only Oxi 340i)
Automatic switchover when the recorder is connected by the cable, AK 323.
mg/l
Concentration
± 0.5 % of the measured value
at an ambient temperature of 5 °C ... 30 °C
%
Saturation
± 0.5 % of the measured value
when measuring in the range of
± 10 K around the calibration temperature
mbar
Partial pressure
± 0.5 % of the measured value
at an ambient temperature of 5 °C ... 30 °C
Temperature
compensation
< 2 % at 0 ... 40 °C
T [°C]
± 0.1
Salinity correction
0 ... 70.0 SAL
Air pressure correction
Automatic through installation of pressure
sensor in the range 500 ... 1100 mbar
Concentration
Voltage
Resolution
0 ... 19.99 mg/l
0 ... 1999 mV
0.01 mg/l per 1 mV
0 ... 90.0 mg/l
0 ... 900 mV
0.1 mg/l per 1 mV
Saturation
Voltage
Resolution
0 ... 199.9
0 ... 1999 mV
0.1 % per 1 mV
0 ... 600 %
0 ... 600 mV
1 % per 1 mV
D. O. partial pressure
Voltage
Resolution
0 ... 199.9 mbar
0 ... 1999 mV
0.1 mbar per 1 mV
0 ... 1250 mbar
0 ... 1250 mV
1 mbar per 1 mV
Accuracy
± 0.5 % of display value ± 0.1 (% saturation)
or
± 0.5 % of the display value ± 0.01 mg/l