Operation
Oxi 330i/340i
20
ba52306e04
03/2008
Note
You can automatically print a calibration protocol after the calibration. To do
so, connect a printer to the interface according to section 4.5.3 before cali-
brating. After a valid calibration, the record is printed.
Sample printout:
Sensor evaluation
After the calibration, the measuring instrument evaluates the current status
of the sensor against the relative slope. The evaluation appears on the dis-
play. The relative slope has no effect on the measuring accuracy. Low values
indicate that the electrolytic filling will soon be depleted and the probe has to
be regenerated.
CALIBRATION PROTOCOL
02.03.99 14:19
Device No.: 12345678
CALIBRATION 02
Cal time: 02.03.99 / 14:19
Cal interval: 14d
OxiCal Tauto AR
Relative Slope: 0,88
Probe: +++
Display
Relative slope
S = 0.8 ... 1.25
S = 0.7 ... 0.8
S = 0.6 ... 0.7
E3
Eliminate the error according to
chapter 6 W
HAT
TO
DO
IF
...
S < 0.6 or S > 1.25