What to do if...
Oxi 330i/340i
48
ba52306e04
03/2008
Error message
E7
AR flashes
continuously
Measured value
too low
Measured value
too high
Display
to
Probe symbol
flashes
Cause
Remedy
Membrane damaged
– Membrane damaged
– Regenerate probe
– Membrane head not screwed
on tight enough
– Screw membrane head tight
Cause
Remedy
No stable measured value
– Membrane contaminated
– Clean membrane
Cause
Remedy
– Insufficient
flow
– Provide flow to the probe
Cause
Remedy
– High amount of dissolved
substances
– Correct solubility function using the
salinity equivalent
– Air bubbles bump on the
membrane with high velocity
– Avoid direct flow to the membrane
– The carbon dioxide pressure is
too high (> 1 bar)
– Measuring not possible
Cause
Remedy
– Time-out of the interface
– Check the instrument that is
connected
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring system