Oxi 330i/340i
Operation
17
ba52306e04
03/2008
4.2.5
Selecting the measuring range, AutoRange
Sensors
You can use the DurOx 325 or CellOx 325 sensors. If the automatic recogni-
tion function
recognizes the DurOx 325 sensor, the measurement is carried
out at a lower resolution. The CellOx 325 sensor measures with a resolution
that is a factor of 10 higher.
AutoRange
Each of the measuring parameters has 2 measuring ranges available.
AutoR-
ange
causes the measuring instrument to change to measuring range II if
measuring range I is exceeded, and also vice versa.
To switch the automatic measurement range selection on/off, see section
4.6 C
ONFIGURATION
. When the AutoRange function is switched on,
ARng
ap-
pears on the display.
Measuring ranges/
resolutions with
CellOx 325 D. O.
probe
Measuring ranges/
resolutions with
DurOx 325 D. O.
probe
%
LoBat
Cal
AR
8
°
F
Sal
mbar
S
1
RCL
Auto Store
1/cm
/
K
%
TDS
UpH
Sal
mV/pH
S/
m
cm
cm
M
Lin
Tref25 Tref20
nLF
AutoCal DIN
AutoCal TEC
Time
Baud
Day.Month No.
Ident
Year
6
7
1
9 2
7
O
°
C
mg/l
ARng
TP
mg/l
%
mbar
Measuring range I
Resolution
0 ... 19.99
0.01
0 ... 199.9
0.1
0 ... 199.9
0.1
Measuring range II
Resolution
0 ... 90.0
0.1
0 ... 600
1
0 ... 1250
1
mg/l
%
mbar
Measuring range I
Resolution
0 ... 19.9
0.1
0 ... 199
1
0 ... 199
1
Measuring range II
Resolution
0 ... 90
1
0 ... 600
1
0 ... 1250
1