Description of the GAMMA FINDER® III
32 / 63
Table 1 (below) shows the energy windows that are set
when the respective nuclide is selected.
Select
Lower energy
threshold [keV]
Upper energy
threshold [keV]
Tc-99m
20
180
I-125
20
60
Co-57
90
150
All
20
512
Use the
º
or
»
button to select the nuclide you want to use
and confirm with the OK button. A confirmation screen will
show you that the corresponding nuclide is now selected.
After a approximately 1.5 seconds or by pressing the OK
button, you return to the OR screen.
5.5.3
Dual Isotope Mode
The Dual Isotope mode can be used to simultaneously
measure the nuclides Tc-99m and I-125 and depict both
counts simultaneously on the display.