Description of the GAMMA FINDER® III
26 / 63
reference measurement is finished. A successful refer-
ence measurement is confirmed with a checkmark and
the measured sensitivity displayed.
The measured sensitivity now serves as the reference value
for all further stability tests.
Performing the regular stability test
To perform a regular stability test, with the GAMMA FIND-
ER® III switched on, press the OK button in the main menu.
Select the item "STABILITY TEST" using the arrow buttons
and confirm with the OK button.
Then select "PERFORM TEST" and confirm with the OK but-
ton.
Place the source holder on the GAMMA FINDER® III so that
the tip of the GAMMA FINDER® III touches the nuclide. Make
sure that the nuclide you used for reference measurement
is inserted in the source holder. The name of the reference
nuclide is displayed in the lower left corner of the screen.