Description of the GAMMA FINDER® III
40 / 63
5.7
Cleaning and disinfecting the GAMMA FINDER® III
The Gamma Finder® III must be wipe cleaned and wipe dis-
infected after each use.
The entire Gamma Finder® III must be wipe cleaned with a
damp cloth in combination with an alcohol-based cleaning
agent.
After cleaning, visually inspect the Gamma Finder® III to
verify that it is free of soil debris or stain. If there are still
visual residuals, repeat the cleaning procedure.
After cleaning, wipe disinfect the entire Gamma Finder® III
with an alcohol-based disinfectant.
WARNING!
Cleaning and disinfection baths
Cleaning and disinfection baths are expressly prohibited.
They can cause serious damage to the GAMMA FINDER® III.