Description of the GAMMA FINDER® III
24 / 63
• Use the OK button to go to the main menu and select
the Stability Test menu item. The Stability Test menu
opens.
• Use the
º
or
»
button to scroll to the "NEW SOURCE"
item and confirm with the OK button.
• Enter the name of the new source. Use the
º
or
»
but-
tons to select the respective letters or numbers. You
confirm a letter or a number with the OK button. Scroll
to the "Finished" item and confirm with the OK button
if the name is entered correctly.
• Then enter the reference date as found in the source
certificate and confirm with the OK button. Pay atten-
tion to the date format.
• Then enter the activity as measured on the reference
date. Scroll to the "Finished" item and confirm with the
OK button.
• Then place the source in the supplied source holder as
depicted in Fig. "Source in source holder" [
25]. Make
sure the writing on the source points to the hole for
the GAMMA FINDER® III and screw down the source
holder.