Description of the GAMMA FINDER® III
29 / 63
After the time has expired, the result of the test will be dis-
played. It is recommended to document the results of the
tests in a continuous table. The test is considered "SUCCESS-
FUL" if the nulleffect of the GAMMA FINDER® III does not ex-
ceed 20 counts per minute (cpm).
Test history
In the "STABILITY TEST" menu, select "TEST HISTORY" to view
the last 10 stability tests performed. Use the
º
or
»
button
to scroll through the various tests. To select details of a test,
press the OK button on the corresponding test.
The details include the source name, the date and time of
the measurement, the age of the source in days, and the
result in cps.
Unsuccessful stability tests
This message indicates that there is no reference measure-
ment and thus no test nuclide has been defined in the
GAMMA FINDER® III. Create a new reference measurement
as described in "Preparing the stability test (reference meas-
urement)".