Description of the GAMMA FINDER® III
36 / 63
5.5.5
Background Suppression
This function facilitates the finding of "99m-Tc enriched" tis-
sue areas, with simultaneously existing relatively high, non-
specific background activity.
To activate the mode, use the OK button to open the main
menu. Then select the "BACKGROUND" item and activate it
with the OK button.
After the function has been triggered, the device calculates
an average of the previous measurements of the non-spe-
cific background activity and from this then sets an activity
threshold. When scanning "hot spots," values exceeding this
threshold will trigger the acoustic signal. The display shows
the detected activity without additional calculations for
control purposes.