Electromagnetic Compatibility
56 / 63
Electromagnetic
interference im-
munity tests
IEC 60601-1-2
test level
Compliance
levels
Electromagnetic
environment/
guidelines
Discharge of
static electricity
(ESD) according
to
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should
be made from
wood or con-
crete or
covered with
ceramic tiles. If
the floor cover-
ing consists of
synthetic ma-
terial, the relat-
ive humidity
should be at
least 30 %.
Electrical fast
transients/
bursts accord-
ing to
IEC 61000-4-4
± 2 kV for
power lines
± 1 kV for input
and output
lines
Repetition fre-
quency 100 kHz
± 2 kV for
power lines
± 1 kV for input
and output
lines
Repetition fre-
quency 100 kHz
The quality of
the supply
voltage should
be the same as
the voltage of a
typical business
or hospital en-
vironment.
Voltage surges
according to
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
The quality of
the supply
voltage should
be the same as
the voltage of a
typical business
or hospital en-
vironment.