Failure

Possible cause

Procedure

Signal span dropping off/too small

Diaphragm is damaged, e.g. through 

impact, abrasive/agressive media; 

corrosion of diaphragm/pressure 

connector; transmission fluid missing.

Contact the manufacturer and replace 

the instrument

Signal span drops off

Moisture present (e.g. at the cable tail) Install the cable correctly

Signal span erratic

Working temperature too high/too low Ensure permissible temperatures as per 

the Operating Instructions

Signal span incorrect

Working temperature too high/too low Ensure permissible temperatures as per 

the Operating Instructions

Abnormal zero point signal

Medium or ambient temperature too 

high/too low

Control the internal temperature of the 

instrument within the permissible range; 

observe the allowable temperature error 

(see Operating Instructions)

Diaphragm is damaged, e.g. through 

impact, abrasive/agressive media; 

corrosion of diaphragm/pressure 

connector.

Replace instrument

Abnormal zero point signal

Working temperature too high/too low Ensure permissible temperatures as per 

the Operating Instructions

Zero point signal unstable/ too low/high Moisture present (e.g. at the cable tail) Install the cable correctly,

filter element insertion

Failure

Possible cause

Procedure

Hot instrument case surface

Permissible ambient or medium tempe-

rature exceeded

Ensure permissible ambient/medium 

temperature limits are observed (see 

Operating Instructions)

9. Trouble shooting

GB

Please verify in advance if pressure is being applied (valves/ ball valve etc. open) and if the 
right voltage supply and the right type of wiring (2-wire) has been chosen? 

9. Trouble shooting / 10. Storage, disposal

GB

If the problem persists, contact our sales department.

USA, Canada

If the problem continues, contact WIKA or an authorized agent for assistance. If the level 
probe must be returned obtain an RMA (return material authorization) number and shipping 
instructions from the place of purchase. Be sure to include detailed information about the 
problem. Level probes received by WIKA without a valid RMA number will not be accepted. 

Process material certificate (Contamination declaration for returned goods)

Purge / clean dismounted instruments before returning them in order to protect our employees 
and the environment from any hazard caused by adherent remaining media. 
Service of instruments can only take place safely when a Product Return Form has been 
submitted and fully filled-in. This Return Form contains information on all materials with which 
the instrument has come into contact, either through installation, test purposes, or cleaning.  
You can find the Product Return Form on our internet site (www.wika.de  / www.wika.com).

In case of unjustified reclamation we charge the reclamation handling expenses. 
*)  Make sure that after the assembly the unit is working properly. In case the error continues to exist send in the instrument for 
 

reparation (or replace the unit). 

10.   Storage, disposal

When storing or disposing of the level probe, take precautions with regard 
to remaining media in removed level probes. We recommend cleaning the 
level probe properly and carefully. Remaining media in the pressure port 
may be hazardous or toxic!

Dispose of instrument components and packaging materials in accordance with the 
respective waste treatment and disposal regulations of the region or country to which    
the instrument is supplied.

Disposal

Mount the protection cap when storing the pressure transmitter in order to prevent any 
damage to the diaphragm.

Storage

Take precautions with regard to remaining media in removed level probes. 
Remaining media in the pressure port may be hazardous or toxic!
Remove the level probe from service and mark it to prevent it from being 
used again accidentally, if it becomes damaged or unsafe for operation.
Have repairs performed by the manufacturer only.

Do not insert any pointed or hard objects into the pressure port for cleaning to prevent 
damage to the diaphragm.

.   Trouble shooting

!

Warning

!

Warning

Содержание IL-10

Страница 1: ...trinsically safe level probe Eigensichere Pegelsonde Sonde de niveau s curit intrins que Sonda de nivel de seguridad intr nseca Current terms and conditions apply Details are available on Es gelten un...

Страница 2: ...in accordance with the intended use IL 10 Use the intrinsically safe level probe to transform the hydrostatic pressure into an electrical signal in hazardous areas 1 Important details for your inform...

Страница 3: ...rec tives of your country and if you have the qualification required You have to be acquainted with the rules and regulations on hazardous areas measurement and control technology and electric circuit...

Страница 4: ...g media in the pressure port may be hazardous or toxic Have repairs performed by the manufacturer only Information about material consistency against corrosion and diffusion can be found in our WIKA H...

Страница 5: ...d in the wall must have the ingress protection according to IEC 60 529 namely for zone 0 IP 67 and for zone 20 IP 6X Integrate an overvoltage protection if the mounting position is less than 1 m away...

Страница 6: ...cal mounting position with lower pressure connection Non linearity of span 0 2 BFSL according to IEC 61298 2 Non repeatability of span 0 1 1 year stability of span 0 2 at reference conditions Permissi...

Страница 7: ...of equip ment intended for use in potenti ally explosive atmospheres 94 9 EC Approval German Lloyd GL Environmental Category C F EMC 1 RF immunity V m 10 Burst kV 4 Wiring protection Short circuit pro...

Страница 8: ...thorized agent for assistance If the level probe must be returned obtain an RMA return material authorization number and shipping instructions from the place of purchase Be sure to include detailed in...

Страница 9: ...f r alle Benutzer jederzeit zug nglichen Ort auf Die nachfolgenden Einbau und Betriebshinweise haben wir mit Sorgfalt zusammengestellt Es ist jedoch nicht m glich alle erdenklichen Anwendungsf lle zu...

Страница 10: ...on 1 Gruppe E F und G Non incendive f r Class I Division 2 Gruppe A B C und D 3 Zeichenerkl rungen Abk rzungen M gliche Gefahr f r Ihr Leben oder schwerer Verletzungen Ex Hinweise M gliche Gefahr f r...

Страница 11: ...eseitigen k nnen Ergreifen Sie Vorsichtsma nahmen f r Messstoffreste in ausgebauten Pegelsonden Messstoffreste k nnen zur Gef hrdung von Menschen Umwelt und Einrichtung f hren Lassen Sie Reparaturen n...

Страница 12: ...Kabeldurchf hrung in der Wand zur Trennung der Zonen muss einer Schutzart nach IEC 60 529 entsprechen und zwar f r die Zone 0 IP 67 f r die Zone 20 IP 6X Integrieren Sie einen berspannungsschutz bei e...

Страница 13: ...ruckanschluss nach unten Nichtlinearit t d Spanne 0 2 BFSL nach IEC 61298 2 Nichtwiederholbarkeit d Spanne 0 1 Stabilit t pro Jahr d Spanne 0 2 bei Referenzbedingungen Zul ssige Temperaturbereiche Mes...

Страница 14: ...wen dung in explosionsgef hrdeten Bereichen 94 9 EG Zulassung German Lloyd GL Environmental Category C F EMC 1 HF Immunit t V m 10 Burst kV 4 Elektrische Schutzarten Kurzschlussfestigkeit Sig gegen UB...

Страница 15: ...10 Lagerung Entsorgung D Im unberechtigtem Reklamationsfall berechnen wir die Reklamationsbearbeitungs Kosten berpr fen Sie nach der Montage die korrekte Arbeitsweise des Systems Besteht der Fehler we...

Страница 16: ...tions 49 9372 132 295 Si le num ro de s rie sur la plaque de fabrication n est plus lisible par exemple par endom magement m canique ou si le num ro est recouvert de peinture la tra abilit n est plus...

Страница 17: ...abr viations 4 Fonction F 2 fils Deux conducteurs servent l alimentation Le courant de l alimentation est le signal de mesure A l aide d un capteur et sous alimentation lectrique on obtient la transf...

Страница 18: ...es de fluides peuvent mettrent en danger les personnes l environnement ainsi que l installation Ne faites effectuer les r parations que par le fabricant Les donn es relatives la r sistance la corrosio...

Страница 19: ...istance entre le point d installation du sonde de niveau et le point de transition de la zone 0 n exc de pas 1 m tre pr voyez une protection contre la surtension Incorporation en zone 0 et zone 20 et...

Страница 20: ...sition verticale raccord de pression vers le bas Non lin arit du gain 0 2 R glage de la plage de tol rance BFSL selon I EC 61298 2 Non r p tabilit du gain 0 1 Stabilit sur un an du gain 0 2 pour les c...

Страница 21: ...rective ATEX pour les appareils tre utilis en atmo sph res explosibles 94 9 EG Homologation German Lloyd GL Environmental Category C F EMC 1 HF Immunit e V m 10 Burst kV 4 Donn es techniques Type IL 1...

Страница 22: ...de s curit pour les restes de fluides se trouvant dans les sondes de niveau d mont s Ces restes de fluides peuvent mettrent en danger les personnes l environnement ainsi que l installation Si vous ne...

Страница 23: ...posible tener en cuenta todos los casos de aplicaci n Si echara en falta informaciones para su problem tica espec fica podr obtener m s informaci n A trav s de nuestra direcci n de Internet www wika...

Страница 24: ...se I Zone 0 AEx ia IIC Dust ignitionproof para clase II y III Division 1 grupo E F y G Non incendive para Class I Division 2 grupo A B C y D 3 Signos s mbolos y abrevaciones Peligro potencial para su...

Страница 25: ...rans misores de presi n desmontados Medios residuales pueden causar da os en personas medio ambiente y equipo S lo el fabricante puede efectuar reparaciones Observe las especificaciones del certificad...

Страница 26: ...o para la separaci n de las zonas debe corresponder a un tipo de protecci n seg n IEC 60 529 o sea para la zona 0 IP 67 y para la zona 20 IP 6X Incorpore una protecci n contra sobretensiones en caso d...

Страница 27: ...de instalaci n conexi n de presi n hacia abajo No Linealidad del rango 0 2 BFSL conforme a IEC 61298 2 No repetibilidad del rango 0 1 Estabilidad al a o del rango 0 2 con condiciones de referencia Ut...

Страница 28: ...ustrial Directiva ATEX sobre los aparatos para uso en atm sfe ras potencialmente explosivas 94 9 EG Homologaci n German Lloyd GL Environmental Category C F EMC 1 Inmunidad radiofrecuencia V m 10 R fag...

Страница 29: ...cuanto a residuos de medios de medici n en transmisores de presi n desmontados Medios residuales pueden causar da os en personas medio ambiente y equipo Ponga fuera de servicio el instrumento y prote...

Страница 30: ...de servicio IL 10 10 Almacenaje eliminaci n de desechos 11 Declaraci n de conformidad CE E Elimine los desechos de componentes de instrumentos y materiales de embalaje seg n el reglamento respectivo d...

Страница 31: ...2146455 08 GB D F E 08 2009 60 WIKA Operating instructions Betriebsanleitung Mode d emploi Instrucciones de servicio IL 10 Original EC declaration of conformity E F...

Отзывы: