EN
48
WIKA operating instructions, model CTR3000
14191761.01 05/2019 EN/DE
6. Commissioning, operation
6.4.5.1 General
After pressing the button
[Mode]
and two options are
visible on the right side: Automatic and manual.
■
Automatic mode requires the setting
Interval
■
Manual mode: A value is taken while pressing the
[Log]
button 3 s long, when being in the application
Home
The interval describes the time when the instrument
takes automatically a value from the measurement.
Clicking on the button on the right side under the
capture
Interval [s]
opens a num pad.
Settable values:
■
Minimum 0.5 s
■
Maximum 3.600 s
■
Step 0.5
■
Standard value = 1 s
The settings must be saved with the button
[Store]
.
Start time describes the time, when the log starts. A
date and a time have to be defined.
The settings must be saved with the button
[Store]
.
Ensure that the date and time is set correct
under
[Settings]
.
Duration time describes the time, when the log ends
after starting. A setting in days, hours or minutes has to
be done.
The settings must be saved with the button
[Store]
.