WIKA operating instructions, model CTR3000
63
14191761.01 05/2019 EN/DE
By selecting the file on the right side, this updates automatically all changes in the probes. Only the lines which are changed
or added are uploaded:
The following main screen shows that there was no update done. Only a new probe was created. Moreover it gives the
information that for 3 probes the coefficients are missing.
6.6 Remote operation
All commands can be found in a separate document.
6.7 Working with a multiplexer
6.7.1 Multiplexer model CTS5000
Multiplexers are detected at power-on and are daisy-chained. Addresses are assigned in order, so the first multiplexer
connected to the CTR3000 will start with the lowest address.
An expansion port connector is provided on the rear panel at the
CTR3000 itself. Here you have to connect the CTS5000 (BUS-in)
with the delivered cable.
The connector is a (1:1) 15-way socket for signal, power and data
connection to the CTS5000-8 or CTS5000-16 multiplexers. It is not
compatible with any other connection.
A 0.5 m [1.64 ft] lead is supplied as standard. Longer lengths may
be used, but system accuracy and the noise specification may be
compromised.
1
1
CTS expansion
Input CTS expansion
6. Commissioning, operation