4. Design and function
4.5.1 Menu selection
4.5.1.1 Applications (apps)
Eight applications are available on the start page:
Home, Settings, Probes, Scan, Logger, Calibration, Remote, Service and Info.
Info is hidden but visible until the menu is scrolled down.
Pos. Application
1
Input title
The input bar is located at the top of the applications.
The menu screen is activated
2
Home application
Use the
[Home]
button to go to the start page.
If the
[Home]
button is held pressed for more than 2.5 seconds, a screenshot with the file name “YYYYMMDD_
hhmmss-Screenshot.png” will be created. This file can be readout via the front-end USB via a memory stick.
3
General settings
Setting or changing of all instrument parameters, such as: Language, date, time, display brightness, temperature
units and temperature probes.
For further information, see chapter 6.4.2 “Application [Settings]”.
4
Probe settings
Create new probes or change the parameters for existing probes e.g. after recalibration.
For further information, see chapter 6.4.3 “Application [Probes]”.
5
Scan settings
Scan function is the sequential measurement of each channel and temporarily shows the data on the display in the
selected view.
For further information, see chapter 6.4.4 “Application [Scan]”.
6
Log
Logger settings
Opens the [Logger] application.
For further information, see chapter 6.4.5 “Application [Logger]”.
7
Cal
Calibration settings
This application allows to calibrate thermometers automatically by changing the set point at user-defined intervals
with a connected dry well CTD9100/9300 or micro calibration bath CTB9100.
For further information, see chapter 6.4.6 “Application [Calibration]”.